Ozone Generator Panel Operation: Start-Up and Shutdown

This guide explains the panel and signal operation of an ozone generator: preparation before start-up, switching on, status indications, automatic operation, and proper shutdown.

1. Preparations before starting the ozone generator

  • Connect the main power supply.
  • Connect the signal lines (used when the user installs a control system or operates the ozone equipment manually).
  • Connect the cooling system (including the user-supplied pressure, volume and flow meter, and shut-off valves).
  • Connect the gas piping (including the pressure regulating valve, flow meter, flow adjustment valve, and shut-off valves).

Note: when the pressure-relief line is closed, pressurized gas must not be fed into the generator, because the safety valve cannot release the pressure; overpressure may damage the equipment.

2. Starting the ozone generator

  • Check that the cooling water volume is within the required range (refer to the technical parameters).
  • Check that the gas pressure and flow are within the required range (refer to the technical parameters).

Once the preparations are complete, release the emergency-stop button. All LEDs will light up and stay lit for a few seconds. If no fault is detected, only the yellow STAND-BY lamp remains lit.

If an ALARM indication appears and does not clear, the corresponding fault message is displayed.

Applying a current pulse to Pin 4 (START) for 0.3 s triggers the high voltage. A delay of 1–2 s is normal. The green OPERATE indicator then lights up and the yellow STAND-BY lamp goes out. If a fault persists, the high voltage will not be switched on.

The control-circuit voltage (current) determines how much voltage is applied to the ozone cell. Operation of the generator is started via the software. When the generator is switched on, the control voltage is already present, and the high voltage ramps up gradually to match the control voltage.

Ozone generator control panel and display states

3. Status indications after start-up

The status of the ozone generator is indicated through relay contacts on the D-SUB signal interface, as follows:

  • Status no. 2 indicates that Pin 1 & Pin 9 are connected.
  • Status no. 3 indicates that Pin 3 & Pin 10 are connected.
  • Status no. 4 indicates that Pin 3 & Pin 10 are connected.
  • Status no. 1 indicates that no external signal is present.

Power setting for remote-control operation

The ozone output power is controlled via a control voltage of 0–5 VDC.

If the generator is controlled with a 0–20 mA current signal, a 250 Ω resistor must be added in the D-SUB plug between the power control input (+) (Pin 14) and the power control input (−) (Pin 7).

Note: the control voltage must stay within 0–5 VDC; otherwise the unit may fault. Be sure to avoid this.

4. Automatic operation

In automatic mode, an external controller regulates the ozone concentration. The measured concentration is obtained from an ozone analyzer such as the BMT 963 or BMT 964. The controller compares the actual value with the setpoint and sends an output signal, which serves as the input signal for the generator’s power control.

The controller output must be within 0–5 V (the input voltage of the ECOPS control), or 0–20 mA (a 250 Ω resistor is required at the ECOPS control input). With this arrangement, a closed-loop control system maintains the preset ozone concentration.

With a SIPART automatic control and calibration unit (Siemens), the generator can be set to run at a fixed output power. In manual mode, the output power of the ozone unit is adjusted with the two arrow keys on the SIPART, over a range of 0–100%.

Compact OZONE CONTROL UNITs (OCU) control all equipment in automatic operation, and additionally integrate mass flow controllers/meters and various electrical interfaces (such as RS232 and PROFIBUS).

If an ozone leak detector shows that the concentration exceeds the TLV (Threshold Limit Value) setpoint, the ozone generator stops working.

5. Switching off the ozone generator

By briefly interrupting the current on the STOP input, the generator switches to STAND-BY mode.

If the ozone output is interrupted for an extended period while the oxygen flow is throttled to a low value, the cooling system also stops, to prevent condensation inside the ozone cell of the unit.

To shut down the unit completely, press the emergency-stop switch to de-energize the generator, and close the oxygen valve and the cooling-system valves for extended shutdown.

If ozone is mixed with water — for example via an injector — the ozone generator should be switched off before the circulation pump of tank 4/9. Under no circumstances should water be allowed to enter the ozone discharge area through the gas system; the user must ensure this by installing suitable check valves and manual valves. If both the ozone unit and the injection water circulation pump are to be shut down, the ozone unit should be switched off first, and the gas line then closed with the manual valves.

Contact Longkang for sizing support and quotations.