UV System Troubleshooting: A Field Checklist

Where lamp diagnosis looks at the tube, this checklist looks at the whole reactor. Most callouts trace back to three symptoms — and most of those trace back to installation, not hardware.
Symptom-to-cause table
| Symptom | Likely cause | Action |
|---|---|---|
| Water at the sleeve end | Cracked quartz; gland not torqued; failed O-ring | Replace sleeve, tighten evenly (do not over-torque), renew seal |
| Low log reduction | Supply voltage low; sleeve fouled; lamp output below 70 µW/cm²; lamp past rated hours; flow above rating; turbidity or suspended solids high | Correct voltage; clean sleeve; replace lamp; check flow; add pre-filtration |
| Lamp will not strike | Blown fuse; loose socket; broken socket tab; failed ballast; failed indicator; damaged lamp | Work from supply to lamp, replacing only what tests bad |
Installation details that prevent the callout
- Vertical mounting: feed low, discharge high, so the chamber stays full and flow slows through the irradiation zone.
- High flow velocity: fit a flow or pressure limiter. Water that rushes past the sleeve does not collect dose.
- Pressure above 3 kgf/cm²: fit a check valve on the outlet to protect the quartz sleeve.
- Solenoid valve: put it on the inlet, upstream of the reactor.
- Turbidity above 5 NTU: pre-filter. Particles shield organisms.
- Service clearance: leave more than 1.2 m at one end for lamp withdrawal and a drain nearby.
Do not skip the safety steps
Never look at a lit lamp directly. Isolate power before opening the chamber, wear UV-rated eye protection and gloves, and let the lamp stabilise for five to ten minutes after striking before judging output. Quartz should be handled with a clean cloth — fingerprints carbonise under UV-C and permanently cut transmission.
One habit pays for itself: log lamp hours, intensity reading and sleeve condition at every visit. A trend tells you whether you are looking at consumable ageing or a design mismatch. Swapping a lamp without logging anything just resets the clock on the same problem.
Need a system sized for your water? Send us your flow rate, UV transmittance (UVT) and the target log reduction, and our engineers will specify the reactor and validated dose. Explore the product range or browse the Technical Library.